T-70 Series Automatic Titrators

Low cost economical and dedicated Titration capability
TA-70 Acidity Titrator, TS-70 Salinity Titrator and the TP-70 Potentiometric Titrator

Additional Information :
T-70 series Brochure (PDF)
Multiple Sample Changer

Send me more:
For more information, quote, etc.

Related:
Process Titrators
SAT-500 Coulometric Chloride Titrator
AUT-701 Automatic titrator

 

Features:
  • Potentiometric single End point Titrations can be performed.
  • LCD with Backlight
  • Results memory for 100 measured samples with re-analysis function of previous data
  • External Printer available for Date, time, mode, parameters, sample No., name, initial and final potential, graph, and results
  • dual RS-232 ports for Sampler interface, External printer, and turntable.
  • Software available for spreadsheet re-analysis
  • Options include a multi-sampler for 12, 24, or 36 samples.

Peripherals and interface software

The T-70 series titrators can accomodate our TTT-510 Sampling device for multiple sample handling. We also have a micro titration adapter for small sample analysis. There is a standard RS-232C interface with software available for donloading insttrument data or real time data to standard spreadsheet format on a PC.

Simple and economical


Each instrument has its own particular method for analysis and calculation capability. They are designater for the type of analysis each instrument is designed for.

Micro-titrations are now possible with the small volume titration adapter.

With the sampling device, you can completely automate your procedure from running the calibration standards, then standardizing the titrant, and then automatically running multiple samples without any operator intervention.

Titration Methods

The T-70 series titrators has the ability to perform the type of titration it is designed for with caclulation capability. This allows for fast and easy set-up for most routine titrations. A recalculation mode allows for re-analysis of previous data.

 

 

AUT-501 printout

External Printer results


To the left is a typical printout which can be obtained with the T-70 series titrators. Notice the analog (or s-shaped) curve as well as the first derivative at the end point.

 

There is software available to output the data results to a PC for use with standard spreadsheet software for further data storage and recall or archiving purposes.

For more information, inquire about our AUT-LOG software control package.

Notice the maximum first derivative value which corresponds to the end point of the titration.

 

Specifications: T-70 series Titrators
Titratior Model TA-70 Acidity Titrator TS-70 Salinity Titrator TP-60 Potentiometric Titrator
Titration Mode 1-Citric Acid, 2-Tartaric Acid, 3-Acetic Aced, 4-Malic Acid, 5-Lactic Acid, 6-Factor mode, 7-Blank Mode 1-Salinity #1, 2-Salinity #2, 3-Factor mode, 4-Blank Mode 1-Inflection Point detection, 2-set point titration mode, 3-pH adjustment mode
Measurement Item/range pH: pH0.00 - 14.00,
mV: 0 to +/-1999.9mV,
Temp: 0 to 100.0 C.
Acidity: 0-100%
mV: 0 to +/-1999.9mV,
Salinity: 0-100% (as NaCl)
pH: pH0.00 - 14.00,
mV: 0 to +/-1999.9mV,
Temp: 0 to 100.0 C.
Conc. Factor: depends on setting
Unit %, ml %, ml %, mg/L, g/L, ml, factor(w), factor(v), blank
Resolution pH: 0.01pH
mV: 0.1mV
Temp: 0.1C
Acidity: 0.001%
mV: 0.1mV
Salinity: 0.001%
pH: 0.01pH
mV: 0.1mV
Temp: 0.1C
Conc: 0.001%
Measuring Range pH: pH0.00 - 14.00,
mV:0 to +/-1999.9mV,
Temp: 0 to 100.0 C.
pH Calibration Automatic 2 or 3 point calibration (pH 4.01, 6.86, 9.18)
Data Memory 100 sets (re-analysis function available)
Display LCD with Back light
Titration Curve, Titration Volume, pH(mV), Temperature simultaneously displayed.
Keyboard Operation Flat Keypad/ Dialogue operation system by function keys and numeric keypad.
Printer External dot matrix Printer is optional. Order EPS-R
Output 2 x RS-232 for Printer, PC or turntable
Burette syringe 20 ml brown glass
Burette Accuracy +/-0.1% whole volume error, Repeatability +/- 0.01ml
Wetted Material Teflon, Daiflon, hard glass.
Power Requirements AC100v-240v, 50/ 60Hz
Power Consumption approx. 33VA
Dimensions approx: 110(w)x396(h)x346(d)mm , approx. 5.6kg

Standard Accessories:

1 ST-7 Magnetic Stirrer
1 01B00001 Electrode Holder
1 6948810K Electrode Stand w/stopper
1 7075610K Unit Base plate (for 1-20ml syringe)
1 7075630K 1L Reagent Bottle, translucent
1 AUT-001P Polyethylene reagent bottle, 500 ml brown
1 7075640K CO2 adsorbent with tube
1 AUT-045P 20 ml Brown glass syringe
1 AUT-066P Metal Plunger puller tool
1 P00070 Brown Glass degassing nozzle (for 1-20ml)
1 AUT-067P Nozzle cleaning wire
1 AUT-022P Teflon Tube, black, 2m
1 P000108 Tube for nozzle purge
1 Manual Operation Manual

Analyticon Instruments Corporation
P.O. Box 92, 500 Morris Avenue
Springfield, NJ 07081 USA
info@analyticon.com
Phone: 973-379-6771 Fax: 973-379-6795