FBM-160 Fluoride Monitor

Industrial Process Fluoride concentration Monitors

 

Additional Information :
FBM-160 / FBM-100A Panel and outdoor enclosure literature (PDF)

Send me more:
For more information, quote, etc.

Related:
FBM-300 Fluoride.monitor (PDF)


General Information

The Models FBM-100A and FBM-160 provide fast and continuous detection of free fluoride ion concentration in water. They are widely used for monitoring water treatment processes and effluent from wastewater plants. They are also used in the semiconductor industry to monitor washed water from plants that use hydrogen fluoride.

The Model FBM-100A is suitable for panel mounting while the Model FBM-160 is designed for outdoor, field mounting. These instuments also feature an optional water jet cleaner for the ion electrode.

The measurement method differs from the more complex JIS K 0102 specified method. It has the advantage of being a much simpler method. However, this measurement method can be influenced by wide pH and temperature variations of the sample. Please refer to the paragraph describing sample conditions to decide on suitability for your particular application.

Features:

  • Fast response time: for a sample containing 2mg/L fluoride ions, an alarm will be generated within 30 seconds at 90% response.
  • Low, middle and high ranges available as standard (0~20, 0~200. 0~2000 mg/L).
  • 4 alarm outputs available. These can be configured based on user’s requirements such as concentration alarms, instrument error, under cleaning, etc.. Concentration alarms are adjustable for delay time and band width.
  • Control outputs (100 VAC) are available for external electrode cleaners such as water-jet cleaner.
  • Output for sample temperature measurement is available (Model FBM-160).
  • RS-232C output signal for measured concentration, sample temperature, concentration alarms etc. is available as an option.

Specifications:
Model Codes: FBM-100A (panel mounting)FBM-160 (outdoor, field mounting)
Measurement Method: Fluoride Ion Selective Electrode
Display: Digital, LCD type
Measurement Ranges: 0.0~99.9 mg/L, 0~999 mg/L or 0~9990 mg/L
Output Signal: 4~20mA DC, isolated, 650 Ohm Load
Output Range: Adjustable within measurement range (minimum 1/10 F.S.). Factory settings: 0.0~20 mg/L, 0~200 mg/L, 0~2000 mg/L.
Sample Temperature: 0~50 deg C
Digital Output Signal (Option): RS232C, Asynchronous, half duplex, 9600 Baud. Data transmitted includes ion concentration, electrode signal, sample temperature, concentration alarms, under maintenance, under cleaning, instrument fault status etc..
Contact Switching Outputs: 4 outputs available (normally open contacts). Select from high limit, low limit, under cleaning, under maintenance, meter error (factory setting is OFF). Delay times and band width are adjustable for the high and low limit alarms. Contact rating: 250 VAC, 3A, or 30VDC, 3A. NOTE: One set of contacts on the Model FBM-160 is normally closed when de-energized. It is possible to use this contact as a power fail alarm output signal.
Cleaning System Control Outputs: Internal timer provides 100VAC, 2A output to drive cleaning system such as water jet cleaner at pre-set intervals. Cleaning cycle adjustable between 0.1~48.0. Cleaning duration adjustable between 1~999 secs. Cleaning pulse adjustable between 1~19 times. Standby time after completion of cleaning adjustable from 0.0 to 99.9 minutes.
Temp Compensation: Fluoride ion electrode is corrected using Nernst equation (within 0~40 deg C of sample temperature).
Performance: Linearity: within +/-8% FS (excluding sensor) within +/-30% FS (with sensor)Repeatability: within +/-5% FS (excluding sensor) within +/-30% FS (with sensor)Response time: (90%): within 15 seconds (excluding sensor) within 60 seconds (with sensor)
Self Diagnostics: Calibration Error: Displays E0~5 Temperature Sensor Error: Displays E-12 Memory Error: Displays E-20/21Burn out or error signal is output
Operating Power: 90~264 VAC, 50/60 Hz
Power Consumption: Approx. 10VA (FBM-100A)Approx. 11VA (FBM-160)

Individual specifications:

FBM-100A

FBM-160

Installation

Panel mounting(panel cut-out : 92 x 92 mm )

Outdoor, filed installation

(50A pipe, wall or rack mounting)

External dimensions

96(w) x 96(h) x 90(d) mm

181(w) x 180(h) x 95(d) mm

Enclosure Rating

Indoor installation type (IP-30)

Outdoor installation type, dust and splash proof (IP-65)

Materials of construction

Main body: AluminiumDisplay: Polyester

Main body: Aluminium die castDisplay: Polyester

Surface finish

Display: Pale Yellow

Main body: Metallic silver

Cable entry

Not applicable

G1/2 x 6 (with 6~12mm diameter cable gland)

Ambient Temp and humidity

-10~50 deg C90% RH or less (no condensation)

-20~55 deg C95% RH or less (no condensation)

Weight

Approx. 0.5 kg

Approx. 2 kg

Water Temp output signal

Not applicable

Adjustable in 10 deg C widths with 1 deg C units. Factory setting 0.0~50.0 degC

 

Principle of Operation

The fluoride ion selective electrode generates an electromotive force corresponding to the fluoride ion concentration in water. The relationship between the concentration and the electromotive force is logarithmically linear as shown in the diagram to the right. The instrument, calibrated with standard solution, can determine the fluoride ion concentration of the sample by just immersing the electrode into the sample.


Sample Conditions

Sample pH:

To be stable within ph4~9 (see Note 1)

Sample Temperature:

To be stable within 0~40 deg C (see Note 2)

Sample Conductivity:

50 mS/m (500 microS/cm) or greater

Flow velocity:

0.01~0.2 m/s

Coexisting components:

Low values of calcium, aluminium, iron etc. (see Note 3)

NOTE 1:This instrument cannot detect fluoride under sample conditions of pH 4 or less because HF exists as molecules (not ionized) under these condition. For samples of pH 9 or above, the influence from OH- ions increases and influences the measurement. Also for large variations in pH, the fluoride conditions change and it may change compounds that cannot be detected. For reliable measurement, the pH value should be kept as stable as possible.

NOTE 2:Calcium salts present in the sample dissolve in greater qualities with increasing sample temperatures causing a positive interference on measurement result. The sample temperature should be kept as stable as possible and kept below 40 degC in order to avoid maintenance troubles.

NOTE3:Calcium, aluminium, iron, etc. can combine with fluoride to form compounds that are different from fluoride ions and cannot therefore be detected by this instrument. Under these circumstances, the measured value would be different to that obtained from the conventional JIS measurement method. The conventional JIS method measures total fluoride (including the above mentioned compounds by distillation method).

 