TOC-200P Portable TOC monitor
TOC-200 Panel transmitter / TOC-1 Detector

Total Organic Carbon in Pure / Ultrapure water, reagent free

Additional Information :

TOC-200 brochure (pdf)

Send me more:
For more information, quote, etc.

Related:

CLF-110/CLF-111 Reagentless Chlorine Monitor

RC-24P Handheld Reagentless Free Chlorine Meter

TOC-200 Portable / Panel mount TOC Analyzer

The DKK-TOA TOC Monitor is used to continuously measure TOC levels in pure water based on the oxidation method using UV light on a TiO2, titanium oxide, catalyst by measuring the electrical conductivity of the water. It also indicates the Conductivity(resistivity) of the water at the same time. The monitor is intended to control process water quality.

Principle of Operation:

The measuring method calculates TOC concentration from the difference before and after oxidation of a sample using a UV light source and TiO2 catalyst. The photo-oxidation catalyst, TiO2, with a high affinity for water, is coated on the internal wall of the oxidizing light source. Water decomposition and OH radical generation takes place on the surface of the catalyst when the UV lamp is activated. The organic carbon component decomposes into CO2 and water. Since CO2 generated in the reaction exists in equilibrium in water, H+ ions and HCO3- ions proportional to the amount of CO2 which will be generated and hence the electrical conductivity of the water increases. We then monitor the increase in conductivity.

The ranges are quite low. 0-20ugC/L, 0-200ug/L and 0-2000ug/L, and dependent on resistivity. The measuring interval is intermittent at 2-5 minute intervals.

It has 4-20 output and RS-232 alarm levels. Flow rate is about 100 mls/minute or less, response is 3 minutes or less for 90% response.

Features:

  • Reagentless operation. No reagent consumption during operation
  • Compact, self contained system. Portable unit is Approx. 340(w)x 280(d)x 435(h) mm and aprox. 7Kgs., IP40 case.
  • Low flow, minimal water consumption: interval measurement: 2-5 minutes, Flow rate: 100ml/minute.
  • High and Low alarm output.
  • Analogue output, 4-20mADC
  • RS-232C output
  • UV lamp life approx. 20,000 hours. Minimal maintenance
  • Response time: approx. 3 minutes (90% response) or less.


Specifications:

Temperature: -5-50C
Consumption: 100ml/min. (intermittent operation, Adjustable 2-5 minutes, variable)
Response time: 90% response within 3 minutes or less (w/flow of 100 ml/min)
Detector Housing: Pyrex glass, quartz glass, TiO2, PPS, PFA, SUS316, FPM and others
Calibration Method by CH3OH reference solution or adjustment to lab analysis
Power Requirements: 85-264VAC ±10%, 50/60Hz.
Power Consumption: Approx. 15-20 VA (max.)
Construction: . Indoor, self-standing or panel mount

 

Range of operation:

The instrument measuring range is according to the table below, three ranges from 0.0 to 20.00ugC/L.

TOC Range
Condition of Resistivity
0.00 to 20.00 ug C/L 10MOhm*cm or more
0.1-200.0 ug C/L 1MOhm*cm or more
0-2000 ug C/L 0.1MOhm*cm or more