HDM-136 / HDM-138 4 wire pH transmitters

Industrial Process pH and ORP Meters

Additional Information :
HDM-136/HDM-138 4 wire Transmitter (PDF)

Send me more:
For more information, quote, etc.

Related:

Portable/handheld pH Meters
Benchtop pH Meters

General Information

The DKK-TOA Industrial process control pH and ORP analyzers and transmitters are available from Analyticon with many options and configurations available to accommodate an entire range of applications. As a leader in analytical sensor technology, many years of expertise and innovation have been incorporated into the finest process control equipment in the world.

Every process monitoring environment will have unique requirements including sensor configuration of the electrode, cell type and configuration, system pressure and cleaning requirements.

In addition to the main controller, Analyticon offers automatic sensor cleaning devices including water jet, ultrasonic, chemical and physical sensor cleaning units.

Industrial pH Control

The HDM-136A series pH analyzers/transmitters are designed as outdoor controllers incorporating a 4 wire signal and AC voltage supply to the unit.

Features:

  • 4 wire unit with built in microprocessor featuring isolated output, digital indication and rainproof construction.
  • Simplified Calibration. Data from up to 5 standards can be stored in memory for single action calibration.
  • Automatic evaluation of electrode quality. Sensor monitors status of electrode during calibration and displays error messages if out of specification.
  • Adjustable pH range
  • Standby mode as "current pH hold" during maintenance work.
  • External output hold command function during cleaning.
  • Measured value shift function for correlation
  • Manual temperature compensation function if needed by electrode system (Non-pt temp probe).
  • Auto determination of 10kOhm and 1kOhm pt temp resistant probe.
  • Automatic temperature compensation for pH. Std conversion at 25C
  • Self Diagnostics. Automatically detects damage to glass membrane, temp probe, abnormal data, and indicated by triggering upper/lower alarm limit (adjustable)

Specifications:
Specification
Product Name

pH Analyzer / transmitter

Model 4-20mADC, Isolated (Standard)
1-5VDC, Isolated
Measuring Range pH: 1.00-14.00 pH
mV: -600-+600mV
Temp. 0-100C, Indication only
Indication Digital LCD display
Resolution pH: 0.01, mV: 0.1
Temp: 0.1C
Performance: Linearity within +/-0.03 pH (at equivalent input)
within +/- 0.06% FS for transmission output
Power requirements

90-132 VAC 50/60Hz
180-264VAC 50/60Hz

Power consumption approx. 3VA
Ambient conditions -20 to 55C, 99%rh or less
Construction Outdoor installation, rainproof type (JIS C0920)
Weight Approx. 3.0kg
Mounting Mounting on 50A pipe
Cable entry G3/4 )PF3/4), 3 ports
Materials

Main: Cast aluminum alloy
Window: Glass

Color Metallic silver and blue
Output signal 4-200mA DC isolated from input. Load resistance: Max 650Ohm
Output range Adjustable in 0.1 pH steps, with minimum width of 2pH

Ordering Information:
Model Order Code Description Specification Notes
HDM-136 1
2
Power Supply

90-132 VAC 50/60Hz
180-264VAC 50/60Hz

AC Power Supply Only
For 24VDC, See model HDM-135
1
2
Transmission Output 4-20mADC, Isolated (Standard)
1-5VDC, Isolated
A
B
C
D
E
F
G
H
J
K
Y
Output Range 0-14 pH (standard)
0-10 pH
0-8 pH
0-14 pH
2-12 pH
3-11 pH
4-14 pH
4-10 pH
6-14 pH
6-12 pH
Special

Use with temp electrodes with 10 kOhm (at 25C)Std temp resistance or 1 kOhm (at 0C), pT resistance probe.

 

Special: output range is also limited.

A
B
Mounting Bracket Provided 304 SUS
Not Provided
 
0
1
2
3
Cable Entrance G 3/4 (PF3/4) (standard)
G 1/2 304 (PF1/2) SUS adapter
NPT 1/2 304 SUS adapter
NPT 3/4 304 SUS adapter
 
0
1
Electrode Lead Connection

Direct Connection
EC-10 Extension cable (for FC-4)

 
0
1
Sunshade Hood Not Provided
Provided 304 SUS
 
A Markings English (others available)  
0 Certification None (others available depending on location)  

 

Industrial ORP Control

The HDM-138A series ORP transmitters are also similar to the HM-136 4 wire series transmitters and designed as outdoor ORP controllers. Refer to above specs. See application sensors for appropriate ORP electrodes which are available.

 

Analyticon Instruments Corporation
P.O. Box 92, 99 Morris Avenue
Springfield, NJ 07081 USA
info@analyticon.com
Phone: 973-379-6771 Fax: 973-379-6795