HBM-160 pH / HBM-162 ORP Wall Mount Monitor

Industrial Process pH and ORP Meters

 

Additional Information :
HBM-160/HBM-162 Enclosure (PDF)

Send me more:
For more information, quote, etc.

Related:
HDM-136/138 pH/ORP transmitter (PDF)
Portable/handheld pH Meters
Benchtop pH Meters

General Information

The DKK-TOA Industrial process control pH and ORP analyzers and transmitters are available from Analyticon with many options and configurations available to accommodate an entire range of applications. As a leader in analytical sensor technology, many years of expertise and innovation have been incorporated into the finest process control equipment in the world.

Every process monitoring environment will have unique requirements including sensor configuration of the electrode, cell type and configuration, system pressure and cleaning requirements.

In addition to the main controller, Analyticon offers automatic sensor cleaning devices including water jet, ultrasonic, chemical and physical sensor cleaning units.

Industrial pH Control

The Model HBM-160 is the pH Monitor and the Model HBM-162 is the ORP Monitor.

 

HBM-160 pH monitor

Features:

  • Microprocessor based 4-wire type transmitter in a robust die-cast aluminum case.
  • Dual output of 4-20mA form pH and temperature.
  • Simplified Calibration. Data from up to 5 standards can be stored in memory for single action calibration.
  • Automatic evaluation of electrode quality. Sensor monitors status of electrode during calibration and displays error messages if out of specification.
  • Automatic and manual temperature compensation functions are available. Electrode use with different temp resistance values is possible.
  • Standby mode as "current pH hold" during maintenance work.
  • External output hold command function during cleaning.
  • Measured value shift function for correlation
  • Dual Temp compensation function annd manual temperature compensation function if needed by electrode system (Non-pt temp probe).
  • Self Diagnostics. Automatically detects damage to glass membrane, temp probe, abnormal data, and indicated by triggering upper/lower alarm limit (adjustable)
  • Dual contact closure outputs for power off and alarm level.
  • Able to accomodate signal output when using sensor washing device (optional), such as BHC-7E / JHC-7E and includes buit in timer and power supply.
  • Operated on AC power, from 90VAC to 264VAC.

Specifications:
Model 160-pH monitor Model 162-ORP monitor
Product Name

Panel/Wall Mount pH Analyzer / transmitter

Panel/Wall Mount ORP Analyzer / transmitter
Model HBM-160 HBM-162
Output signals

4-20mADC, Isolated (Standard), Max load 650 Ohms
Optional: 4-20mA with RS-232C

4-20mADC, Isolated (Standard), Max load 650 Ohms
Optional: 4-20mA with RS-232C
Measuring Range pH: -1.00-15.00 pH
Electrode EMF: -800-+800mV
Temp. -5 to-100C, Indication only
ORP: -2000 to +2000 pH
Electrode EMF: -800-+800mV
Temp. -5 to 100C, Indication only
Indication Digital LCD display Digital LCD display
Resolution 0.01 pH: , 1 mV
Temp: 0.1C
0.01 pH: , 1 mV
Temp: 0.1C
Performance: Linearity: within +/-0.03 pH (with equivalent input)
Repeatability: Within +/-0.02pH
Linearity: within +/- 4mV (with equivalent input)
Repeatability: Within +/- 3mV
Power requirements

90-264VAC 50/60Hz

90-264VAC 50/60Hz
Power consumption approx. 10VA approx. 10VA
Ambient conditions -20 to 55C, 95%rh or less -20 to 55C, 95%rh or less
Construction Outdoor installation, rainproof type (IP65 equivalent) Outdoor installation, rainproof type (IP65 equivalent)
Weight Approx. 2.0kg Approx. 2.0kg
Mounting Mounting on 50A pipe (option: wall rack) Mounting on 50A pipe (option: wall rack)
Cable entry 6 cable glands for 6-12mm cable, removable for flex conduit(G1/2 x 6) 6 cable glands for 6-12mm cable, removable for flex conduit(G1/2 x 6)
Materials

Main: die cast aluminum alloy
Window: Polyester

Main: die cast aluminum alloy
Window: Polyester
Color Metallic silver Metallic silver
Control Outputs

2 Contacts: 250VAC, 3A or less. functions include sensitivity 0.01 to 2.0pH, 0-60 sec delay, etc.

Option: 1, 3 or 4 contact outputs

2 Contacts: 250VAC, 3A or less. functions include sensitivity 1 to 200mV, 0-60 sec delay, etc.

Option: 1, 3 or 4 contact outputs

Output range width 2 pH or more, freely set at 0.01pH steps width 400 mV or more, freely set at 1mV steps
Self Diagnosis During Measurement: cracks in glass pH membrane, electrolyte shortage, temp comp malfunction.
During chemical cleaning (RHC-7E/HCl): electrode response.
During calibration: seven types of diagnostic information are displayed.
During Measurement: electrolyte shortage, temp comp malfunction.
During chemical cleaning (RHC-7E/HCl): electrode response.
Sensor information

Use Model HC-7 or HC-8 electrode holders
Compatible electrodes include Model 5600, 5610, etc.
Connector box FC-4, Cable EC-10

Use Model HC-7 or HC-8 electrode holders
Compatible electrodes include Model 2600, 2610, etc.
Connector box FC-4, Cable EC-10

Ordering Information:
Model 160 Order Code Description Specification Notes
HBM-160-0- 1

Power Supply

90-264VAC 50/60Hz

The voltage ordered for main unit is also passed through to washing device. Be sure voltages are compatible when ordering.
1
2
Transmission Output 4-20mADC, Isolated (Standard)
4-20mADC and RS-232C interface
A
B
C
D
E
F
G
H
J
K
Y

pH Transmission Output Range
(-1.00-15.00pH)

0-14 pH (standard)
0-10 pH
0-8 pH
0-14 pH
2-12 pH
3-11 pH
4-14 pH
4-10 pH
6-14 pH
6-12 pH
Special, Other spec.

Use with temp electrodes with 10 kOhm (at 25C)Std temp resistance or 1 kOhm (at 0C), pT resistance probe.

 

Special: output range is also limited.

A
Y

Liquid Temp. Output range
(-5-100C)
0-50C
Other Specification(10C or more in width, units of 1C
0
1
2
Alarm Output 2 point alarm (2 circuit with contact closure)
2 point alarm (2 circuit with contact and power off signal)
4 point alarm(3 circuit with contact a, 1 with c)
1
2
Washing timer Not included
Included(for JHC-7E, RHC-7EC, PHC-7E, etc.
When using "E" series together, select 2 . With timer not included, Transmission output hold function operates based on contact closure from washing device that has a timer function
1
2
Mounting Bracket 50A pipe, 304 SUS
wall mount
0
1
2
Cable Entrance Standard (Cable gland for 6-12mm cable, 6 ports)
G 1/2 (RF1/2)
NPT 1/2
0
1
Air purge No
Yes
0
1
Sunshade Hood Not Provided
Provided- 304 SUS
0
1
Markings/indication None
English (others available)
0
1
Certification None
Inspection certificate (others available depending on location)

 

Industrial ORP Control

The HBM-162 Series ORP monitor Specifications
Model 160 Order Code Description Specification Notes
HBM-162-0- 1

Power Supply

90-264VAC 50/60Hz

The voltage ordered for main unit is also passed through to washing device. Be sure voltages are compatible when ordering.
1
2
Transmission Output 4-20mADC, Isolated (Standard)
4-20mADC and RS-232C interface
A
B
C
D
E
F
G
H

Y

ORP Transmission Output Range
(-2000-+2000mV)

-200-+200mV
-700-+700mV
-1000-+1000mV
-1400-+1400mV
-2000-+2000mV
-300-+700mV
0-+1000mV
0-+1400mV

Special, Other spec.

Use with temp electrodes with 10 kOhm (at 25C)Std temp resistance or 1 kOhm (at 0C), pT resistance probe.

 

Special: output range 400mV or more, units of 10mV.

A
Y

Liquid Temp. Output range
(-5-100C)
0-50C
Other Specification(10C or more in width, units of 1C
0
1
2
Alarm Output 2 point alarm (2 circuit with contact closure)
2 point alarm (2 circuit with contact and power off signal)
4 point alarm(3 circuit with contact a, 1 with c)
1
2
Washing timer Not included
Included(for JHC-7E, RHC-7EC, PHC-7E, etc.
When using "E" series together, select 2 . With timer not included, Transmission output hold function operates based on contact closure from washing device that has a timer function
1
2
Mounting Bracket 50A pipe, 304 SUS
wall mount
0
1
2
Cable Entrance Standard (Cable gland for 6-12mm cable, 6 ports)
G 1/2 (RF1/2)
NPT 1/2
0
1
Air purge No
Yes
0
1
Sunshade Hood Not Provided
Provided- 304 SUS
0
1
Markings/indication None
English (others available)
0
1
Certification None
Inspection certificate (others available depending on location)

 

Analyticon Instruments Corporation
P.O. Box 92, 99 Morris Avenue
Springfield, NJ 07081 USA
info@analyticon.com
Phone: 973-379-6771 Fax: 973-379-6795

Copyright 1995-2005, legal and trademark. all rights reserved. v.01/2005