General Information
The DKK-TOA Industrial
process control pH and ORP analyzers and transmitters are available from
Analyticon with many options and configurations available to accommodate
an entire range of applications. As a leader in analytical sensor technology,
many years of expertise and innovation have been incorporated into the
finest process control equipment in the world. 
Every process monitoring
environment will have unique requirements including sensor configuration
of the electrode, cell type and configuration, system pressure and cleaning
requirements.
In addition to the
main controller, Analyticon offers automatic sensor cleaning devices including
water jet, ultrasonic, chemical and physical sensor cleaning units.
Industrial pH Control
The Model HBM-160
is the pH Monitor and the Model HBM-162 is the ORP Monitor.
HBM-160 pH monitor
Features:
- Microprocessor
based 4-wire type transmitter in a robust die-cast aluminum case.
- Dual output of
4-20mA form pH and temperature.
- Simplified Calibration.
Data from up to 5 standards can be stored in memory for single action
calibration.
- Automatic evaluation
of electrode quality. Sensor monitors status of electrode during calibration
and displays error messages if out of specification.
- Automatic and manual
temperature compensation functions are available. Electrode use with
different temp resistance values is possible.
- Standby mode as
"current pH hold" during maintenance work.
- External output
hold command function during cleaning.
- Measured value
shift function for correlation
- Dual Temp compensation
function annd manual temperature compensation function if needed by
electrode system (Non-pt temp probe).
- Self Diagnostics.
Automatically detects damage to glass membrane, temp probe, abnormal
data, and indicated by triggering upper/lower alarm limit (adjustable)
- Dual contact closure
outputs for power off and alarm level.
- Able to accomodate
signal output when using sensor washing device (optional), such as BHC-7E
/ JHC-7E and includes buit in timer and power supply.
- Operated on AC
power, from 90VAC to 264VAC.
Specifications:
|
Model
160-pH monitor |
Model
162-ORP monitor |
| Product
Name |
Panel/Wall Mount
pH Analyzer / transmitter
|
Panel/Wall
Mount ORP Analyzer / transmitter |
| Model |
HBM-160 |
HBM-162 |
| Output
signals |
4-20mADC, Isolated
(Standard), Max load 650 Ohms
Optional: 4-20mA
with RS-232C
|
4-20mADC,
Isolated (Standard), Max load 650 Ohms
Optional: 4-20mA
with RS-232C |
| Measuring
Range |
pH:
-1.00-15.00 pH
Electrode EMF: -800-+800mV
Temp. -5 to-100C, Indication only |
ORP:
-2000 to +2000 pH
Electrode EMF: -800-+800mV
Temp. -5 to 100C, Indication only |
| Indication |
Digital
LCD display |
Digital
LCD display |
| Resolution |
0.01
pH: , 1 mV
Temp: 0.1C |
0.01
pH: , 1 mV
Temp: 0.1C |
| Performance:
|
Linearity:
within +/-0.03 pH (with equivalent input)
Repeatability: Within +/-0.02pH |
Linearity:
within +/- 4mV (with equivalent input)
Repeatability: Within +/- 3mV |
| Power
requirements |
90-264VAC 50/60Hz
|
90-264VAC
50/60Hz |
| Power
consumption |
approx.
10VA |
approx.
10VA |
| Ambient
conditions |
-20
to 55C, 95%rh or less |
-20
to 55C, 95%rh or less |
| Construction |
Outdoor
installation, rainproof type (IP65 equivalent) |
Outdoor
installation, rainproof type (IP65 equivalent) |
| Weight |
Approx.
2.0kg |
Approx.
2.0kg |
| Mounting |
Mounting
on 50A pipe (option: wall rack) |
Mounting
on 50A pipe (option: wall rack) |
| Cable
entry |
6 cable glands for 6-12mm cable, removable for flex conduit(G1/2 x
6) |
6
cable glands for 6-12mm cable, removable for flex conduit(G1/2 x 6) |
| Materials |
Main: die cast
aluminum alloy
Window: Polyester
|
Main:
die cast aluminum alloy
Window: Polyester |
| Color |
Metallic
silver |
Metallic
silver |
| Control
Outputs |
2 Contacts:
250VAC, 3A or less. functions include sensitivity 0.01 to 2.0pH,
0-60 sec delay, etc.
Option: 1, 3
or 4 contact outputs
|
2 Contacts:
250VAC, 3A or less. functions include sensitivity 1 to 200mV, 0-60
sec delay, etc.
Option: 1, 3
or 4 contact outputs
|
| Output
range |
width
2 pH or more, freely set at 0.01pH steps |
width
400 mV or more, freely set at 1mV steps |
| Self
Diagnosis |
During
Measurement: cracks in glass pH membrane, electrolyte shortage, temp
comp malfunction.
During chemical cleaning (RHC-7E/HCl): electrode response.
During calibration: seven types of diagnostic information are displayed. |
During
Measurement: electrolyte shortage, temp comp malfunction.
During chemical cleaning (RHC-7E/HCl): electrode response.
|
| Sensor
information |
Use Model HC-7
or HC-8 electrode holders
Compatible electrodes include Model 5600, 5610, etc.
Connector box FC-4, Cable EC-10
|
Use
Model HC-7 or HC-8 electrode holders
Compatible electrodes include Model 2600, 2610, etc.
Connector box FC-4, Cable EC-10 |
Ordering
Information:
| Model
160 |
Order
Code |
Description |
Specification |
Notes |
| HBM-160-0- |
1
|
Power
Supply |
90-264VAC 50/60Hz
|
The
voltage ordered for main unit is also passed through to washing device.
Be sure voltages are compatible when ordering. |
|
1
2 |
Transmission
Output |
4-20mADC,
Isolated (Standard)
4-20mADC and RS-232C interface |
|
|
A
B
C
D
E
F
G
H
J
K
Y |
pH Transmission
Output Range
(-1.00-15.00pH)
|
0-14
pH (standard)
0-10 pH
0-8 pH
0-14 pH
2-12 pH
3-11 pH
4-14 pH
4-10 pH
6-14 pH
6-12 pH
Special, Other spec. |
Use with temp
electrodes with 10 kOhm (at 25C)Std temp resistance or 1 kOhm (at
0C), pT resistance probe.
Special: output
range is also limited.
|
|
A
Y
|
Liquid
Temp. Output range
(-5-100C) |
0-50C
Other Specification(10C or more in width, units of 1C |
|
|
0
1
2
|
Alarm
Output |
2
point alarm (2 circuit with contact closure)
2 point alarm (2 circuit with contact and power off signal)
4 point alarm(3 circuit with contact a, 1 with c) |
|
|
1
2 |
Washing
timer |
Not
included
Included(for JHC-7E, RHC-7EC, PHC-7E, etc. |
When
using "E" series together, select 2 . With timer not included,
Transmission output hold function operates based on contact closure
from washing device that has a timer function |
|
1
2 |
Mounting
Bracket |
50A
pipe, 304 SUS
wall mount |
|
|
0
1
2
|
Cable
Entrance |
Standard
(Cable gland for 6-12mm cable, 6 ports)
G 1/2 (RF1/2)
NPT 1/2
|
|
|
0
1 |
Air
purge |
No
Yes |
|
|
0
1 |
Sunshade
Hood |
Not
Provided
Provided- 304 SUS |
|
|
0
1 |
Markings/indication |
None
English (others available) |
|
|
0
1 |
Certification |
None
Inspection certificate (others available depending on location) |
|
Industrial ORP
Control
The HBM-162 Series
ORP monitor
Specifications
| Model
160 |
Order
Code |
Description |
Specification |
Notes |
| HBM-162-0- |
1
|
Power
Supply |
90-264VAC 50/60Hz
|
The
voltage ordered for main unit is also passed through to washing device.
Be sure voltages are compatible when ordering. |
|
1
2 |
Transmission
Output |
4-20mADC,
Isolated (Standard)
4-20mADC and RS-232C interface |
|
|
A
B
C
D
E
F
G
H
Y |
ORP Transmission
Output Range
(-2000-+2000mV)
|
-200-+200mV
-700-+700mV
-1000-+1000mV
-1400-+1400mV
-2000-+2000mV
-300-+700mV
0-+1000mV
0-+1400mV
Special, Other spec. |
Use with temp
electrodes with 10 kOhm (at 25C)Std temp resistance or 1 kOhm (at
0C), pT resistance probe.
Special: output
range 400mV or more, units of 10mV.
|
|
A
Y
|
Liquid
Temp. Output range
(-5-100C) |
0-50C
Other Specification(10C or more in width, units of 1C |
|
|
0
1
2
|
Alarm
Output |
2
point alarm (2 circuit with contact closure)
2 point alarm (2 circuit with contact and power off signal)
4 point alarm(3 circuit with contact a, 1 with c) |
|
|
1
2 |
Washing
timer |
Not
included
Included(for JHC-7E, RHC-7EC, PHC-7E, etc. |
When
using "E" series together, select 2 . With timer not included,
Transmission output hold function operates based on contact closure
from washing device that has a timer function |
|
1
2 |
Mounting
Bracket |
50A
pipe, 304 SUS
wall mount |
|
|
0
1
2
|
Cable
Entrance |
Standard
(Cable gland for 6-12mm cable, 6 ports)
G 1/2 (RF1/2)
NPT 1/2
|
|
|
0
1 |
Air
purge |
No
Yes |
|
|
0
1 |
Sunshade
Hood |
Not
Provided
Provided- 304 SUS |
|
|
0
1 |
Markings/indication |
None
English (others available) |
|
|
0
1 |
Certification |
None
Inspection certificate (others available depending on location) |
|
|